American Journal of Epidemiology Vol. 120, No. 4: 542-549
Copyright © 1984 by The Johns Hopkins University School of Hygiene and Public Health
research-article |
A DEVICE FOR THE AUTOMATIC MEASUREMENT OF BLOOD PRESSURE IN EPIDEMIOLOGIC STUDIES
1Departments of Medicine and Pediatrics, University of Massachusetts Medical School Worcester, MA
2Department of Cardiology, Children's Hospital, and Department of Pediatrics, Harvard Medical School Boston, MA
3Science Department, The Phillips Exeter Academy Exeter, NH
Reprint requests to Dr. Ellison, Division of General Medicine, University of Massachusetts Medical Center, 55 Lake Avenue North, Worcester, MA 01605
An automatic blood pressure apparatus was devised for use in an epidemiologic study that required self-administered measurements of blood pressure for up to 200 students each week over a school year. A Dinamap device was used to determine systolic and diastolic blood pressures by the osciliometric method. An Apple II Plus computer with one disc drive was used for control of the device and storage of data. Two such apparatuses were assembled and used to make more than 6,000 recordings of blood pressure over a six-month period. Except for some initial problems with a computer component, the apparatuses performed well; successful blood pressure recordings were ob tained 99% of the time. Use of the automatic devices proved to be very acceptable to students. It is concluded that the use of such devices offers a reasonable means for obtaining accurate measurements of blood pressure and may be applicable for many epidemiologic studies.
blood pressure determination; computers; oscillometry
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
K. M. Rose, D. K. Arnett, R. C. Ellison, and G. Heiss Skip Patterns in DINAMAP-Measured Blood Pressure in 3 Epidemiological Studies Hypertension, May 1, 2000; 35(5): 1032 - 1036. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. A. Mundt, L. E. Chambless, C. B. Burnham, and G. Heiss Measuring Ankle Systolic Blood Pressure: Validation of the Dinamap 1846 SX Angiology, July 1, 1992; 43(7): 555 - 566. [Abstract] [PDF] |
||||
![]() |
M. W. Gillman, S. A. Oliveria, L. L. Moore, and R. C. Ellison Inverse Association of Dietary Calcium With Systolic Blood Pressure in Young Children JAMA, May 6, 1992; 267(17): 2340 - 2343. [Abstract] [PDF] |
||||
![]() |
R. C. Ellison, A. L. Capper, R. J. Goldberg, J. C. Witschi, and F. J. Stare The Environmental Component: Changing School Food Service to Promote Cardiovascular Health Health Educ Behav, January 1, 1989; 16(2): 285 - 297. [PDF] |
||||